Risk Control Certifies Rates, Not Magnitudes: An Audit of Selective Fetal Biometry
Zhengshuai Yang, Linna Jiang, Changyun Li, Guihu Zhao · 4 Nov 2026 · arXiv
Abstract
Prenatal ultrasound biometry must return a head circumference (HC) and refuse when the image cannot support one. Selective prediction trades coverage against error, but its finite-sample guarantees bound the rate of gross errors, not their magnitude. We audit a tool-grounded biometry pipeline on two separately pre-registered four-way splits of the 999 labelled HC18 images, retraining the segmenter three times on each. Calibrating the contour-completeness threshold by exact Clopper-Pearson risk control returns a certificate indistinguishable from a hand-set floor, yet the pipeline still emits a 271.3 mm error: a small, geometrically complete ring fitted to an image whose true HC is 324.6 mm. The accepted candidate scores 0.94 completeness, so the score cannot see the failure, and one outlier among $\approx 150$ answers leaves the certified risk bound ($α= 0.10$) intact. A second, untouched split reproduces the mechanism on a different image (233.3 mm). Constraining the admissibility region---the anatomical range in which the measurand is defined---removes the family instead: raising the lower bound from 50 to 55 mm costs 0.4 coverage points and cuts the worst-case error on both spli